csm_Pour_Point_Process_Analyzer_615d95a697csm_Pour_Point_Process_Analyzer_615d95a697

Pour Point Process Analyzer

PPA-4
  • ASTM compliant measurement based on tilting mechanism
  • Low and high temperature applications
  • Opacity independent measurement
  • Network and fieldbus communication
Description
Technical Data
Questionnaires/Requirement Sheets
Data Sheets
Brochures

To remain competitive, today’s refiners must employ all optimization and product control techniques available. The use of online physical property analyzers is one of the key features to reach those objectives because they measure important quality properties in the process directly.The pour point of a liquid is the temperature at which it still flows but starts losing its flow characteristics by becoming semi solid. For hydrocarbons the pour point temperature depends on the content of paraffin in the liquid but also on the viscosity that changes with temperature. The pour point temperature is an important quality parameter especially for lube oils but also for gas oils and fuel oils.

Marking

ATEX: II 2G Ex h IIC T4 Gb XIECEx: Ex IIC T4 GbNEC 500: Class I, Division 2, Group B,C and DNEC 505: Class I, Zone 1, AEx d e ib px IIBresp. IIB+H2 T3 resp. T4CEC Sec. 18: Ex d e ib px IIC T3 resp. T4TR CU: II Gb T4 X

Ambient Temperature

T4

Technical Data

Technology automatic tilting measuring cell
Method compliant with: ASTM D97, DIN EN ISO 3016, IP 15 correlates with: ASTM D5949 Automatic Tilt Method
similar to ASTM D5950
Sample at inlet filtered 50 μm, free of suspended water (≤ 37 cSt at inlet temperature)
approx. 20 to 40 l/h
1
to 3 bar (14.5 to 43.5 psi)
normal: 30 to 50 °C (86 to 133 °F) min. 20 K above pour point
temperature
Sample at outlet atmospheric pressure
Electrical data max. 8 (4 to 20 mA; 1000 Ω); active isolated on request
4 to 20 mA; 160 Ω
24 V DC; max. 0.5
A
high: 15 to 28 V DC low: 0 to 4 V DC
24 V DC; max. 0.8 A
Communication MODBUS/RTU via RS485 or RS422 or FOC is
MODBUS/TCP via FOC is
pour point temperature (others on
request)
Alarm, Ready/Valid
Stream Selection, Validation Request, Reset
via Ethernet (VDSL
or FOC is)
Ambient conditions operation 5 to 40 °C (41 to 104 °F) storage -20 to 60 °C (32 to 140 °F)+
operation 5 to 80 %
relative humidity, non-corrosive; storage 5 to 80 % relative humidity, non-corrosive
IP 54
(comparable with NEMA 13)
HMI touch display 19″
remote access on request
cyber security
operating system Windows10
enterprise LTSB
Bartec Benke control software PACS
Process connections tube fittings: typically Swagelok 6mm/12mm/18mm; others on request
sample inlet up to 2 sample
streams and 1 validation stream
sample outlet
vent line (if required)
coolant (if required)
Weight and dimensions approx. 300 kg (without options)
approx. 1140 x 1900 x 710 mm
free space right: 150 mm/left: 100
mm

Questionnaires/Requirement Sheets

    Questionnaire_PL AMS_PAT

Data Sheets

    BENKE Pour Point Process Analyzer PPA-4

Brochures

    BENKE Pour Point Process Analyzer PPA-4
    Cyber Security Solutions for BARTEC BENKE analyzers